7 - 1
Siemens AG
RX57-029.061.01
Page 1 of 4
MEMOSKOP 2/3/50/100/SUB
Medical Engineering
Rev. 02
10.97
TD SD 24
Laser camera timing (not for MEMOSKOP 2)
7
Timing diagram
7
This timing diagram is for reference only.
Changing the times
7
Prerequisites
7
Input can only be done if there is a MEMOSKOP keyboard on hand.
The values will only be displayed if there is a text module. However, "blind" programming
can also be done.
For UROSKOP and Multiline, plug D1.X6 on the MEMOSKOP must be removed when
programming the times.
Opening Setup
7
•
Press the [NEXT PAT] key.
•
Next, press the [@] and [%] keys in sequence .
•
Press the [CTRL] + [S] keys simultaneously.
•
A list of numbers will be displayed on the monitor.
Changing the precontact time TV
7
•
Enter the following : 00006506
**
**
means in HEX code: 01 = 100 ms, 02 = 200 ms, 32 = 5000 ms.
•
Then press the [CTRL] + [E] keys simultaneously.
•
Then, depending on the requirement, program the contact time TK and/or postcontact
time TN or exit the Hardcopy Setup (refer to the section "Exit Setup ").
Adhere to the manufacturer’s instructions for the particular
camera being used.
TV
TK
TN
TV = Precontact time default = 100 ms
TK = Contact time, default = 200 ms
TN = Postcontact time default = 5000
Monitor B, BAS
, Image ON
Image OFF
Hardcopy, BAS
, Image ON
Image OFF
Relay contact ON
Relay contact OFF
Notice