30
Installing the system components
MAMMOMAT Novation
SPB7-250.812.03.02.02
Siemens AG
08.04
CS SD 24
Page 30 of 36
Medical Solutions
Fig. 22: Screws for the base plate
Pos. 1
Leveling screw
Pos. 2
Cover plate
Pos. 3
Floor plate
Pos. 4
Caulk all around using sanitary silicon
Pos. 5
E.g. Hilti heavy duty anachor bolts HSL M8/20
NOTICE
The screws must not come in contact with any conductive materi-
als such as reinforcing iron, since the grounding could be compro-
mised.
¹
Stay clear of conductive materials when fastening the
screws.