Installing the system components
23
Page 23 of 36
Medical Solutions
Installing the cable ducts
0
1. Position the cable ducts
to the stand and the radiation shield respec-
tively.
2. Leave sufficient space, approx. 10 mm, between the cable ducts
and
the base plate of the stand
to allow for the mounting of the cable out-
let cover
.
3. Check that the cable duct
is slightly overlapped by the cable outlet
cover
.
Fig. 13: Installing the cable ducts (top view)
1. Mark the outlines of the cable ducts
on the floor.
2. Cut the bottom of the cable ducts
and the cable duct cover
to proper length.
NOTE
Installing the cable ducts
shows a right-angled in-
stallation of the cable ducts.
If the console is to be installed at another angle to the stand, you
just have to cut the cable ducts correspondingly.
3. Fasten the bottom plate of the cable ducts
to the floor according to the
marking using double-sided adhesive tape.
4. At the radiation shield, cover the cable entry not used with the enclosed cover
plate
.