Area dose product measurement device
5
Siemens AG
SPR2-240.814.04
Page 5 of 10 POWERMOBIL/ARCOSKOP
Medical Engineering
Rev. 02
07.00
TD SD 24
Calibrating the area dose product measuring device
1
CONSTANT parameters = T
w
1
•
Read and record the
T
w
value shown on the electronics unit of the area dose product
measuring device.
Programming
1
•
Connect the service PC to the serial interface of the POWERMOBIL/ ARCOSKOP.
•
Start the service program.
•
Select the "Adjustment, Parameter..." menu.
•
Select "Diamentor" in the "Parameter Groups" combo box.
•
Click on "Get from Unit".
•
The parameter "1. Diamentor Existing" must be on YES.
•
Select parameter "2. Constant
T
w
.
•
Enter the parameter
T
w
in the "Value Actual:" field.
•
Click on "Set Value".
•
Click on "Put to Unit".
•
Click on "Save to File".
Fig. 1
This adjustment is used to calibrate the area dose product mea-
suring device to the POWERMOBIL/ ARCOSKOP display.
You must exit the service program with "Logoff" and "Quit" in
order to save the modified parameters to diskette as a new file.
Potentiometer
NOTICE
NOTICE