Area dose product measurement device
3
Siemens AG
SPR2-240.814.04
Page 3 of 10 POWERMOBIL/ARCOSKOP
Medical Engineering
Rev. 02
07.00
TD SD 24
Installation
1
•
Disconnect the POWERMOBIL respectively ARCOSKOP from the mains power.
•
Remove the POWERPHOS cover.
Without the laser targeting device:
1
•
Attach the Area dose product measurement device to the POWERPHOS using the
screws included in the packaging.
•
Connect the interface cable already integrated in the cable harness to the Area dose
product measurement device and lock it in place.
•
Reattach the POWERPHOS cover.
With the laser targeting device:
1
•
Remove the laser targeting device from the POWERPHOS by removing the screws.
•
Mount the Area dose product measurement device to the POWERPHOS.
•
Reattach the laser targeting device.
•
Connect the interface cable already integrated in the cable harness to the Area dose
product measurement device and lock it in place.
•
Check the laser targeting device according to the adjustment instructions and,
if necessary, adjust it.
•
Reattach the POWERPHOS cover.