POWERMOBIL/ARCOSKOP

 

SPR2-240.814.04

Page 4 of 10

Siemens AG

 

Rev. 02

07.00

TD SD 24

Medical Engineering

4

Area dose product measurement device

 

Memoskop Programming

1

Turn on the POWERMOBIL/ARCOSKOP and wait for the start-up.

Insert the Memoskop parameter disk (see System folder Register 10) in drive A of the 
service PC.

Start the Memoskop service program.

Click the "Get from Memoskop" button in the main menu. The parameters are transferred 
from the Memoskop to the Service PC.

Select the "Others" menu.

Select "Disable dose display selection in User Set Up". The Check button is not marked.

Select "Storage of dose on disk". The Check button is marked.

Click on the "Back" button. The "Others" window closes.

Select the "Hardcopy Setup" menu.

Select "Dose on Hardcopy output". The Check button is marked.

Click on the "Back" button. The "Hardcopy Setup" window closes.

Click on the "Put to Memoskop" button in the main menu. The parameters are 
transferred from the service PC to the Memoskop and saved.

Select the "File" menu.

Select the sub-menu "Save" and save the parameters on disk.

Exit the Memoskop service program.

Turn the POWERMOBIL/ARCOSKOP off and the on again to reset the serial interface.

After selecting the "Disable dose display selection in User Set Up" 
parameter, the "Storage of dose on disk" and "Dose on hardcopy output" 
parameters become available in the Memoskop User Setup and visible to 
the customer.

NOTICE

Summary of Contents for ARCOSKOP

Page 1: ...ction transmission or use of this document or its contents is not permitted without express written authority Offenders will be liable for damages All rights including rights created by patent grant or registration of a utility model _or_ design _are_ reserved SP Installation and Setting Instructions Area dose product measurement device ...

Page 2: ...ments required 1 POWERMOBIL respectively ARCOSKOP system binder including the general safety information and the service instructions If there is a laser targeting device Setting and adjustment instructions for the Laser Targeting Device Parts required 1 Measurement and auxiliary devices required 1 Tools required 1 Standard service kit refer to ARTD Part 3 1 set of Allen keys Installation kit Area...

Page 3: ...ace cable already integrated in the cable harness to the Area dose product measurement device and lock it in place Reattach the POWERPHOS cover With the laser targeting device 1 Remove the laser targeting device from the POWERPHOS by removing the screws Mount the Area dose product measurement device to the POWERPHOS Reattach the laser targeting device Connect the interface cable already integrated...

Page 4: ...rage of dose on disk The Check button is marked Click on the Back button The Others window closes Select the Hardcopy Setup menu Select Dose on Hardcopy output The Check button is marked Click on the Back button The Hardcopy Setup window closes Click on the Put to Memoskop button in the main menu The parameters are transferred from the service PC to the Memoskop and saved Select the File menu Sele...

Page 5: ...COSKOP Start the service program Select the Adjustment Parameter menu Select Diamentor in the Parameter Groups combo box Click on Get from Unit The parameter 1 Diamentor Existing must be on YES Select parameter 2 Constant Tw Enter the parameter Tw in the Value Actual field Click on Set Value Click on Put to Unit Click on Save to File Fig 1 This adjustment is used to calibrate the area dose product...

Page 6: ...ue is displayed on the monitor The tested Tw value is displayed on the monitor of the POWERMOBIL ARCOSKOP This value must match the programmed CONSTANT parameter Tw Tolerance 5 If the 0 key is pressed reset key for fluoroscopic time the test is repeated and is displayed again on the monitor The potentiometer Fig 2 of the area dose product measuring device can be used to adjust the test value Tw to...

Page 7: ... chamber in the center of the I I Fig 4 Dose measurement 1 Press the Next Pat key to reset the area dose product display on the POWERMOBIL ARCOSKOP monitor to 0 Select kV STOP and set 70 kV Reset the dose measuring device Dali Nomex to 0 Release fluoroscopy for approx 20 seconds 3mGy The measured dose KE should be high enough that the selected measurement range of the dose measuring device Dali No...

Page 8: ...vice Evaluation 1 Calculate the deviation between the measured area dose product and the displayed area dose product The deviation must be 0 3 Deviation displayed area dose product FDPa measured area dose product FDPg measured area dose product FDPg Fig 3 Fig 4 Fig 5 Measuring chamber POWERPHOS RBV I I Centering cross POWERPHOS RBV I I ...

Page 9: ...RCOSKOP Medical Engineering Rev 02 07 00 TD SD 24 Concluding steps 1 Reinstall all the covers Perform the protective conductor measurement as specified in ARTD 002 731 17 Perform partial acceptance tests in compliance with country specific regulations Changes to previous version 1 Page 4 new ...

Page 10: ...POWERMOBIL ARCOSKOP SPR2 240 814 04 Page 10 of 10 Siemens AG Rev 02 07 00 TD SD 24 Medical Engineering 10 Area dose product measurement device This page intentionally left blank ...

Reviews: