POWERMOBIL/ARCOSKOP
SPR2-240.814.04
Page 2 of 10
Siemens AG
Rev. 02
07.00
TD SD 24
Medical Engineering
2
Area dose product measurement device
1
Safety information
1
When performing service work and tests, adhere to the general safety instructions, print
number RA0-000.12.40 contained in Register 1 of the system binder, the product-specific
safety information in the POWERMOBIL respectively ARCOSKOP as well as the safety
information in the ARTD. After completing all service work and reattaching the covers,
perform the protective conductor measurement as specified in ARTD-002.731.17.
Documents required
1
POWERMOBIL respectively ARCOSKOP system binder including the general safety
information and the service instructions.
If there is a laser targeting device: Setting and adjustment instructions for the Laser
Targeting Device.
Parts required
1
Measurement and auxiliary devices required
1
Tools required
1
Standard service kit (refer to ARTD, Part 3)
1 set of Allen keys
Installation kit: Area dose product measurement device
Service PC
Service PC-Host interface cable
99 00 440 RE999
POWERMOBIL resp. ARCOSKOP
Protective conductor test meter
44 15 899 RV090
or
Bender Safety tester
PTW-DALI dose measurement device
(no longer listed in the ARTD)
or
PTW-NOMEX dose measurement device
97 08 637 Y0388
or
PTW-DIADOS dose measurement device
97 17 612 Y0388
Center cross
96 60 051 RE999