11.1 CellaView via USB point-to-point connection
Install the USB driver
Connect the pyrometer to the PC
Start CellaView
Select the correct COM port or use the CellaView search function.
For more information on how to work with CellaView read the separate
CellaView manual.
11.2 CellaView via RS485 point-to-point connection
A USB/RS485 converter is needed when working with the RS485 inter-
face. The pyrometer can be directly connected if the PC has an integrat-
ed RS485 Interface board. The distance between pyrometer and PC can
be up to 1200 m.
In order to prevent reflections on the RS485 connection, it is absolutely
necessary to terminate the connection at the pyrometer and at the con-
verter. The termination is integrated in the pyrometer and is active when
delivered from the factory.
Moreover, use a converter with galvanic isolation (e.g. W&T 38211) to
avoid problems with ground loops.
Summary of Contents for 7MC3040-...
Page 1: ...Digital Radiation Pyrometer PA 1x 2x 3x 7MC3040 Operating Manual 04 2017 ...
Page 2: ......
Page 8: ...s Operating manual PA ...
Page 93: ...s Operating manual PA 85 29 Dimensions ...
Page 97: ...s Operating manual PA 89 31 1 Mounting angle PA 11 K Pyrometer PA Mounting angle PA 11 K ...
Page 98: ...s Operating manual PA 90 31 2 Polarising filter Pyrometer PA Polarising filter PA 20 P ...
Page 101: ...s Operating manual PA 93 31 6 Cable VK 02 A Ident Nr 101 3909 ...
Page 102: ...s Operating manual PA 94 31 7 Cable VK 02 F Ident Nr 103 1446 ...
Page 103: ...s Operating manual PA 95 32 Mounting combinations 32 1 Mounting combination PA 20 007 ...
Page 104: ...s Operating manual PA 96 32 2 Mounting combination PA 20 010 ...
Page 111: ...s Operating manual PA 103 ...
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......