User’s responsibility
The pyrometer may only be used when it is in perfect working condition.
3.3
Safety requirements
The instrument works with an operating voltage of 24 VDC. The voltage re-
quired for operation must be supplied by a separate power supply. This
power supply unit must conform to directive DIN IEC 61010.
3.4
Electromagnetic Compatibility
The devices comply with the essential safety requirements of the Elec-
tromagnetic Compatibility Directive 2014/30/EU (EMC Act).
When connecting a power supply unit, make sure that is also conforms
to these standards. Radio interference may arise if the pyrometer is in-
terconnected with such peripheral devices which have not been properly
interference-suppressed. This may necessitate additional interference
suppression measures.
4 General Description
The ARDOCELL PA Series provides efficient, microprocessor-controlled
pyrometers for non-contact temperature measurements
The field of applications for PA 1x includes temperature detection of ma-
terials such as synthetic materials, rubber, textiles, paper, coated sheet
steel, wood or varnish within a range from 0 °C to 1000 °C.
The ARDOCELL PA 13 is specially designed to detect the temperature
of objects in flame-heated furnaces. Because PA 13 works at a narrow
spectral range
of 3.9 μm, water vapour and CO
2
in
the detector’s field of
view will not influence the measurement, even at greater target distanc-
es. The spectral sensitivity of this model enables highly accurate and re-
liable measurements through flames and combustion gases. This py-
rometer can also tackle other atmospheric challenges such as reducing
the effect of background radiation within.
The ARDOCELL PA 15 was specially developed for temperature meas-
urements of glass. Glass is transparent in the visible spectral range and
in the near infrared range. Its emissivity depends on the wavelength, the
type and the thickness of the glass. In the range of 4.5 - 8 µm glass has
an emissivity of almost 100%. Above 5 µm, atmospheric influences, such
as humidity or water vapour, affect the measurement. The ARDOCELL
PA 15 is provided with a blocking filter with a spectral sensitivity of 4.46 -
Summary of Contents for 7MC3040-...
Page 1: ...Digital Radiation Pyrometer PA 1x 2x 3x 7MC3040 Operating Manual 04 2017 ...
Page 2: ......
Page 8: ...s Operating manual PA ...
Page 93: ...s Operating manual PA 85 29 Dimensions ...
Page 97: ...s Operating manual PA 89 31 1 Mounting angle PA 11 K Pyrometer PA Mounting angle PA 11 K ...
Page 98: ...s Operating manual PA 90 31 2 Polarising filter Pyrometer PA Polarising filter PA 20 P ...
Page 101: ...s Operating manual PA 93 31 6 Cable VK 02 A Ident Nr 101 3909 ...
Page 102: ...s Operating manual PA 94 31 7 Cable VK 02 F Ident Nr 103 1446 ...
Page 103: ...s Operating manual PA 95 32 Mounting combinations 32 1 Mounting combination PA 20 007 ...
Page 104: ...s Operating manual PA 96 32 2 Mounting combination PA 20 010 ...
Page 111: ...s Operating manual PA 103 ...
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......