POLYMOBIL Plus
Medical Solutions
Rev. 02
03.05
CS PS 24
When working with the system open, there is danger of
! Electric shock !
•
The capacitor bank can still be charged.
Do not attempt to work on the system while this condition exists.
•
After switching off the system, approximately 450 VDC (350 VDC
for Serial-No. < 10100) may still be present in the system; even
after disconnecting the line voltage plug. Within 10 minutes this
voltage will drop to approximately 10 V.
•
Always measure the actual voltage present with the DVM at test
points -VCC and +VCC on the D960 inverter board or (more
accessible) on the D970 capacitor board at the + connection
point of capacitor C3 and on the right side of fuse F3 (refer to
Fig. 1).
•
LED’s V1 ... V10 on D 970 go out at a significantly higher voltage
level and therefore they are not reliable safety indicators.
•
If a fuse on the D 970 has responded, high voltage may still be
present at the affected capacitor even after a prolonged period
of time.
•
The capacitor discharging circuit utilizes the D 925 board and
relays CS and LS. If connector X3 or X9 on D 925 or D 950 are not
inserted or there is a defect in the circuit, the C-bank will not
discharge.
This can cause life-threatening voltage to be present in the
system even after a prolonged period of time.
•
Refer also to the POLYMOBIL Plus Service instructions
RXR8-120.061... "Replacing the capacitor bank".