Electrical safety / reports
Measuring circuit
Direct measurement
.
Differential measurement
Fig. 2
Measuring circuit for direct measurement of the device leakage
current according to DIN VDE 0751-1:2001-10, Fig. C5
for protective class
I
.
1 = System
2 = Application part (if available)
3 = Measuring arrangement (integrated in the measuring
instrument)
Fig. 3
Measuring circuit for measuring the device leakage current
according to the differential current method
according toDIN VDE 0751-1:2001-10, Fig. C6
for protective class
I
.
1 = System
2 = Application part (if available)
3 = Measuring arrangement (integrated in the measuring
instrument)