Optional Accessories -
LSB-Moduel with 4 Digital Outputs
91
9.3.4 Limit Value Monitoring (LI)
Exceeding/Undershooting the Limit Value Display
See section 5.5
Parameter 930 / 940 / 950 / 960
Selection of the variable to be monitored for limit value 1 (2, 3, 4)
0 = off, 1 = measured O
2
value, 2-7 = configurable measurement value 1-7, 8 =
temperature probe, 9 = absolute pressure probe, 10 = probe current, 11= probe
voltage
Parameter 931/932 / 941/942 / 951/952 / 961/962
Form for maximum comparison value for LI 1 (2, 3, 4)
0 = off, 1 = constant value, 2-13 = calculated analog value 1-12
Form for minimum comparison value for LI 1 (2, 3, 4)
0 = off, 1 = constant value, 2-13 = calculated analog value 1-12
Parameter 933 / 943 / 953 / 963
Constant for maximum comparison value for LI 1 (2, 3, 4)
(only when 931, 941, 951, 961= constant value)
Parameter 934 / 944 / 954 / 964
Constant for minimum comparison value for LI 1 (2, 3, 4)
(only when 932, 942, 952, 961= constant value)
Parameter 935 / 945 / 955 / 965
Reset mode for limit value 1 (2, 3, 4)
0 = automatic, 1 = manual, 2 = acknowledge
Parameter 936 / 946 / 956 / 966
Trigger delay for limit value 1 (2, 3, 4)