Fig. 47: Separation of firmware into an application part and a metrologically relevant part
Legend
1
Metrologically relevant part
2
Application
•
The firmware is separated into an
application part
and a
metrologically relevant part
.
•
The separation allows customer-specific parameters and application software data to
be changed via the SOPAS configuration software without violating the LFT conformity
of the multi-dimensional measurement system.
Fig. 48: Calibration switch for interlocking the metrological software parameters
•
The software relevant to the LFT conformity with the accompanying LFT parameters is
protected by a calibration switch.
•
The calibration switch is located beneath the sealed service flap.
•
It is interlocked by turning the calibration switch to the zero position.
Firmware
Calibration switch