8.4.3
Teaching in analog outputs
8.4.3.1
Teaching in the measuring range
1.
Push the piston to the start of the measuring range.
2.
Press the keypad.
3.
Hold the keypad for less than 2 seconds.
✓
The yellow LED flashes slowly.
2 Hz
4.
Lift your finger.
✓
The first measuring point is defined.
5.
Push the piston to the end of the measuring range.
6.
Press the keypad.
✓
The measuring range has been defined.
NOTE
The process of teaching in the measuring range may be aborted (yellow LED flashes 4
times):
•
No magnet detected when teaching in the start or end of the measuring range.
•
Value below minimum measuring range (> 1 mm).
8
OPERATION
24
O P E R A T I N G I N S T R U C T I O N S | MPS-C
8019721.1FIH/2022-06-01 | SICK
Subject to change without notice