![SICK LD-MRS400001 Operating Instructions Manual Download Page 83](http://html1.mh-extra.com/html/sick/ld-mrs400001/ld-mrs400001_operating-instructions-manual_1264621083.webp)
Operating Instructions
Chapter
6
LD-MRS
Commissioning and configuration
8012948/ZMQ0/2017-06-09
©
SICK AG · Germany · All rights reserved · Subject to change without notice
83
6.3.2
Start the SOPAS-ET and open the scanning assistant
1. Start the SOPAS-ET.
SOPAS-ET opens the program window with the English user interface by default.
2. To change the language setting, click C
ANCEL
in the start dialogue and change the user
interface language to an other language in the menu by going to T
OOLS
, O
PTIONS
.
3. Once the language has been changed, close SOPAS-ET and restart it.
4. In the dialog window, select the option C
REATE
NEW
PROJECT
and confirm with OK.
5. In the main window under S
CAN
ASSISTANT
, click the button C
ONFIGURATION
.
The S
CAN
ASSISTANT
dialog appears.
6.3.3
Configure the ethernet connection
Important
Deactivate all the programs on your computer that access the ethernet or TCP/IP.
1. In the S
CAN
ASSISTANT
dialogue
,
go to I
NTERNET
PROTOCOL
, IP
COMMUNICATION
and check the
box A
CTIVATE
IP
COMMUNICATION
and U
SE
A
UTO
IP.
2. Confirm the settings in the S
CAN
ASSISTANT
dialog by clicking OK.
The S
CAN
ASSISTANT
dialog closes.
6.3.4
Run a scan
1. In the S
CAN
ASSISTANT
dialogue, click the button S
TART
SCAN
.
2. Select the listed devices and confirm with A
DD
DEVICE
.
The connected devices are searched over the connection.
SOPAS-ET adds the listed devices to the project tree.
6.4
Initial startup
The LD-MRS is adapted to the local field monitoring by SOPAS-ET. A user-specific set of pa-
rameters is created with SOPAS-ET. The parameters set can either be loaded from the de-
vice (upload), or it can be created independently.
The parameter set is then loaded onto the LD-MRS (download). This is either done immedi-
ately (the SOPAS-ET option D
OWNLOAD
IMMEDIATELY
) or manually (the SOPAS-ET command
DOWNLOAD
ALL
PARAMETERS
TO
THE
DEVICE
).
Important
After configuration, permanently save the parameter set on the LD-MRS.