DATOSPIR TOUCH User’s Manual 63
511-B00-MU2 • REV. 1 .09
A)
FLEISCH PNEUMOTACHOMETER
1
Remove the filter and then the
transducer by pressing its tab so
that it comes away from its
fixtures. See the figure.
2
Immerse the transducer and filter in a
CIDEX® OPA
solution (follow the
manufacturer's instructions). Then rinse them
in destilled water.
Do not use abrasive substances or solvents.
3
Shake the transducer and filter for removing the water droplets,
leave them to dry at room temperature and re-assemble the
group again.
B)
TURBINE TRANSDUCER
1
Remove the turbine from its housing by
pressing slightly its tab so that it comes
away from its fixtures.
2
Immerse the turbine in a
CIDEX® OPA
solution (follow the
manufacturer's instructions). Then rinse the turbine by
immersion in destilled water. Do not rinse the turbine by holding
it under running water.
Do not use abrasive substances or solvents.
3
Given that the turbine reliability depends on its good condition,
examine it for possible damages.
4
Leave it to dry at room temperature and re-assemble the turbine
in the housing.