The data screen shows the observed values in PRE and POST mode
and the method of comparison between them, depending on the
option selected in the
CUSTOMIZATION MENU
.
Weighted % between PRE and POST
% between REF (Predicted) and POST
% between PRE and POST
Difference between PRE and POST
Thus, it may provide a single report of spirometric results before
(PRE) and after (POST) application of a bronchodilator drug.
4.4 SLOW VITAL CAPACITY «VC» TEST
1
Access the test by pressing
, from the main screen.
•
The axes are always displayed in VOLUME/TIME mode.
•
The equipment saves a maximum of 8 maneuvers ordered
according to the VC value, where M1 is the best VC and M8 the
worst.
•
The maximum time allowed for the maneuver is 45 seconds.
•
To measure the ERV and TV parameters correctly, each
maneuver must have at least four respiratory cycles.