DATOSPIR TOUCH User’s Manual 37
511-B00-MU2 • REV. 1 .09
3
Enter the patient’s details by pressing on each field.
An alphanumeric or numeric keyboard that will appear,
depending on the field data type required. Enter patient’s data
and press
. If patient already exists in the database, press
to retrieve the patient’s details from memory.
Once patient selected, the patient’s ID will appear at the top of
the screen. When finish press
to go to the test screen.
To achieve high quality results, always use the nose
clip.
Always use bacterial filters when you suspect of patient
infection.
4
Instruct the patient on the test performance, his cooperation is
essential for proper implementation.
The test must be carried out by medical staff. Review spirometry
technique bibliography or request information about spirometry
courses to SIBEL S.A.U.
5
When the patient is ready and have assumed a correct posture,
fit the nose clip. The subject must hold the transducer, without
moving it, and wait until a blinking arrow appears on the screen.
The transducer must remain in the same position until the
end of the maneuver.
Correct
Incorrect