38
-
4 Previous Data (Measurement Value)
When "4 Previous Data" is selected in the
main menu display, the file names of all data
obtained by past measurements including the
latest (previous) measurement data are
displayed in chronological order.
are
displayed when there are more file names that
cannot fit on one screen.
The file names of measurement data show the date/time that sampling was started. The
first file name "0406301510" in the above example indicates the file obtained by sampling
that started at "2004, June 30, 15 hours 10 minutes." When the measurement data file is
selected by pressing the [UP] and [DOWN] keys, and the [ENTER] key is pressed, the
measurement data can be checked in the same way as for the Last Data menu. (
For
details, see P29.)
Press the [MENU/ESCAPE] key several times according to the screen display until the
display returns to the main screen.
*
The maximum number of measurement data files that can be displayed in the Previous
Data menu is 99. When sampling is performed with the sampling count exceeding 99,
data is not logged in chronological order at 1-minute intervals. When the sampling count
has reached 99, clear memory (Clear Logs).
5 Clear Logs (Erase Memory)
When "5 Clear Logs" is selected in the main
menu display, the screen on the right is
displayed. If the [ENTER] key is pressed in
this state, all data obtained by past
measurements including the latest
measurement data will be cleared from
memory.
While data is being cleared, "Data erasing"
will be displayed for about five seconds. When memory is cleared, "Completed" is
displayed and the screen returns to the main menu display.
*
If the [MENU/ESCAPE] key is pressed before the [ENTER] key is pressed in the Clear
Logs screen display, the screen will return to the main menu display without memory
being cleared. Press the [MENU/ESCAPE] key in the main menu display to return to
the main screen.
*
If the [POWER] switch or a key on the panel is operated or the power is turned OFF
while data is being cleared (while "Data erasing" is displayed), memory will be
damaged, and recording of measurement data might no longer be possible.
*
When memory becomes full of data, data may no longer be logged in chronological
order at 1-minute intervals and power may no longer be recovered.
*
Even in the event of a malfunction or other problem, Sibata shall not provide
compensation for any data or records that the user failed to obtain or lost, or for any
other direct or indirect damage or losses related to that data. Therefore, please
periodically back up all data as a precaution in case of a malfunction or accident. It is
recommended that functionality be verified in advance and that the instrument be
periodically inspected.
P r e v i o u s
D a t a
1
: 0 4 0 6 3 0 1 5 1 0
▼
2
: 0 4 0 7 2 0 0 8 2 2
3
: 0 4 0 8 1 5 0 7 3 0
4
: 0 4 0 9 3 0 0 9 5 3
5
: 0 4 1 0 2 0 1 0 0 0
6
: 0 4 1 1 1 0 1 3 1 0
7
: 0 4 1 2 0 1 1 6 1 5
C l e a r
L o g s
P r e s s
[ E N T E R ]
t o
e r a s e
l a s t
a n d
a l l
p r e v i o u s
d a t a .