3 Setup (Settings)
When "3 Setup" is selected in the main menu, the following sub-menu is displayed. Select
the menu in the same way as for the main menu. Press the [MENU/ESCAPE] key several
times according to the screen display until the display returns to the main screen.
Setup Menu Display
Display Description
1 Set Flow Rate
Sets the suction flow rate.
2 Set ON Time
Sets the ON timer.
3 Set OFF Time
Sets the OFF timer.
4 Set Date & Time
Sets the clock.
5 Select Timer
Selects the timer.
6 Set Flow Unit
Sets the flow rate unit.
7 Set Fault Stop
Sets the fault stop.
Setup – Set Flow Rate (Suction Flow Rate Setting)
When "1 Set Flow Rate" is selected and the
[ENTER] key is pressed, the preset flow
rate is displayed. The setting range for the
suction flow rate is 10.0 to 20.0 L/min. The
default is 16.7 L/min. When the numerical
value is changed by the [UP] and [DOWN]
keys and the [ENTER] key is pressed, the
newly set value is entered, and the screen
returns to the Setup menu.
* After setting a new value, if the [MENU/ESCAPE] key is pressed before the [ENTER]
key is pressed, the screen will return to the Setup menu without the newly entered value
being accepted.
Setup - Set ON Time (ON Timer Setting)
When "2 Set ON Time" is selected and the [ENTER] key is pressed, the Set ON Time
menu is displayed. Select the respective menu by the [UP] and [DOWN] keys, and press
the [ENTER] key.
Set ON Time Menu Display
Display Description
1 Delay Time
Sets the delay time timer.
2 Start Time
Sets the start time timer.
Delay Time (Delay Time Timer Setting)
When "1 Delay Time" is selected and the
[ENTER] key is pressed, the set value of
the sampling start delay time is displayed.
The setting range of the delay time is 0.00
to 999.59 (hours.minutes). The default is
0.00.
After the "hours (hrs)" setting is changed by the [UP] and [DOWN] keys and the
[ENTER] key is pressed, the "minutes (min)" setting can be changed. Likewise, when
the "minutes" setting is changed in the same way as "hours" and the [ENTER] key is
pressed, the newly set values are entered. The screen returns to the Set ON Time
menu.
S e t u p
S e t
F l o w
R a t e
L / m i n
A c t u a l
V F
1 6.7
S e t u p
O N
T i m e
S e t
D e l a y
T i m
e
h r s
m i n
0 0
2 4.