21
4.
Calibration
4.1.
Precautions and Use
Calibration is the process to ensure the device will scan with the optimal accuracy and scan quality.
Each EinScan 3d scanner was factory-tested and achieved, at least, the listed accuracy from the
technical specification according to VDI/VDE 2634 standards.
Besides factory tested guarantee of accuracy, to assure high scanning accuracy, each EinScan is also
coming with a calibration board for customer calibration. This customer calibration is not required
daily often. Only when the first time after scanner and software are installed. Or, when you are in the
following situations, you should calibrate the device:
◆
Device change
◆
After device enduring bumpy transportation
◆
After device accuracy decreases
◆
When data misaligned very often
4.2.
Preparation for Calibration
Take out the calibration board, and the position paper.
Calibration board
Position paper
Note:
1.
Customer calibration is recommended to do twice a week for daily use.
2.
Make sure to protect the calibration board and keep it clean, no scratches or stains on the black
surface with white circles.
3.
The Calibration board is matched to the Device with same Serial Number. Doing the calibration
with an incorrect calibration board will fail to generate good scan data or optimum accuracy.
4.
For common use, you are not suggested to wipe calibration board. If you still need to clean it,
clean with clear water only and do not use chemical liquid (alcohol, for example).