26
5. Communication and Other Settings
5.2 Settings
There are 8 DIP switches numbered from 1 to 8. 5 items including 3 communication settings can be
set by DIP switches. Settings are listed in the following.
* indicates the default settings.
5.2.1 Communication Setting 1 (Baud rate)
DIP switch number
1
2
3
Baud rate
OFF OFF OFF
300
ON OFF OFF
600
OFF* ON* OFF*
1200*
ON ON OFF
2400
OFF OFF ON 4800
ON OFF ON
9600
OFF ON ON
19200
Settings
ON ON ON
38400
5.2.2 Communication Setting 2 (Parity)
DIP switch number
4
5
Parity
OFF* OFF*
No*
ON OFF
No
OFF ON
Even
Settings
ON ON
Odd
5.2.3 Communication Setting 3 (Delimiter)
DIP switch number
6
Delimiter
OFF* CR*
Setting
ON LF