![Shimadzu EP-90 Instruction Manual Download Page 24](http://html1.mh-extra.com/html/shimadzu/ep-90/ep-90_instruction-manual_1256947024.webp)
19
4. Functions and Operation
4.8 Auto Print Mode
Auto print function prints out the weighed result automatically as a sample is placed on the balance
pan and the display stabilizes. Productivity is enhanced as the operator doesn't have to press any key.
(1) Press A.P. key
. The auto print function of the balance is activated and Auto Print lamp
of EP-90 will be illuminated.
(2) Press TARE key or O/T key of the balance or EP-90. The balance display will be zeroed.
(3) Place the sample on the balance pan. The weighed result will be automatically printed out
when the display stabilizes.
(4) Press A.P. key
while Auto Print lamp is illuminated to turn off the auto print function.
Notes
Conditions for data output in auto print mode vary depending on the balance series.
Refer to the balance instruction manual for UW/UX series, BW/BX series, BW-K/BX-K
series and AUW-D/AUW/AUX/AUY series. A.P. key of EP-90 only turns it ON/OFF. With
balance series for which detailed conditions of auto print function can be selected
(UW/UX, BW-K/BX-K etc.), the current settings on the balance remain valid. Refer to the
balance’s instruction manual for the details. Auto print is activated even with the balance
series for which auto print cannot be set on the balance (BL, EL/ELB etc.). Upon stability
is resumed after sample is loaded at zero display, the mass value is automatically
printed out.
4.9 Date
Output
Press DATE key
. Command is sent to the balance to receive date and time from the balance’s
built-in clock. Date and time are then printed out using one line each.
Every time DATE key
is pressed, DATE lamp
will be turned on/off. While DATE lamp is
on, date and time are printed for each mass value that is printed.
Note
Date output function of EP-90 is not available with a balance that is not equipped with
built-in clock.