47
6
Trouble
Shooting
Other Messages
The following message is displayed.
Radiography is not possible while this message is displayed.
Action
Message
Description
Door/Interlock
The examination room door
is open.
Close the door.
FPD (Water) Flow Error
Heat Exchanger of FPD
detects error.
Contact Shimadzu service
representative.
(It's possible to perform
fluoroscopy and
radiography)
C-arm
■
Action
Message
Description
Reset with the STOP
switch or turn off and on
the power again. If the
equipment cannot be
reset, contact Shimadzu
service representative.
Motor Error XXX
Motor command fail to drive
the motor.
Camera Rotation Error
Error of FPD rotation
potentiometer.
Collimator Error XXX
Error of the collimator
potentiometer.
Collimator Rot Error
Error of collimator rotation
potentiometer.
Limit Error XXX
Detects the limit switch
activation.
Ask Shimadzu service
representative to readjust
the limit switch.
It is possible to operate
the C-arm by holding
down the release switch.
Reset with the STOP
switch or turn off and on
the power again. If the
equipment cannot be
reset, contact Shimadzu
service representative.
Position Detect Error XXX
Mismatching of position
detection and motor drive.
Potentio Error XXX
Detects disconnection of
position detector circuit.