46
Messages Related to the X-ray Tube
The following message is displayed.
Radiography is not possible while this message is displayed.
Action
Message
Description
Stop the operation until
the heat unit value
reduces.
Thermal Over
HU Over Predicted
HU Full Stored
Value predicted to go
out of range Continuing
radiography with present
setting will cause heat unit
overrun.
X-ray tube unit temperature
exceeded the permitted limit.
Heat units reached the
permitted limit.
(LX-Tube Water) Flow Error
Water flow in heat
exchanger of X-ray tube is
stopped.
Contact Shimadzu service
representative.
(LX-Tube) Water Level Warning
Water volume of heat
exchanger gets low
representative.
Contact Shimadzu service
representative.
(It is possible to perform
radiography though this
warning message is
displayed)
Error Messages at Power ON
Contact Shimadzu service
representative.
Panel Battery Warning
The data recording battery in
the operation panel must be
replaced soon.
Cabinet Battery Empty
The data recording battery in
the control cabinet is empty.
Cabinet Battery Warning
The data recording battery in
the control cabinet must be
replaced soon.
Panel Battery Empty
The data recording battery in
the operation panel is empty.
Action
Message
Description