18
Check the following items before using the system.
1
Before turning on the power switch. Ensure there is:
•
No damage on the exterior.
•
No Condensation or Water on the system.
•
No twisting of the cables.
2
After turning on the power switch. Ensure there are:
•
No abnormal sounds.
•
No abnormal smells.
•
No error messages on X-ray generator console, Acquisition monitor or Review monitor.
3
Check the operation of the C-arm and Catheterization Table.
Confirm that each operation functions correctly and there are no abnormal sounds.
•
LAO-RAO / CAUD-CRAN operation of the C-arm.
•
Change the position of the C-arm and move the FPD forward/backward.
•
Up-and-down operation of Catheterization Table and Tabletop Lock Switches.
4
Check the Safety Switch operation.
•
Actuation of the Safety Switch of the Collimator and FPD.
•
Operation of Stop Switches.
5
Confirm the Image Field Size Selector.
•
While performing fluoroscopy, change and confirm the field of view size of the FPD.
6
Confirm the operation of the Collimator.
•
Check to be sure that the shutter of the Collimator is moving relative to the FPD field of
view change.
7
Check to be sure that images are displayed.
•
Make sure that fluoroscopy/radiography images are displayed on monitors properly.
Daily Inspection