13. Application Functions
13.7 Formulation Mode
This function is convenient for making many measurements of minute samples and seeking the total
mass.
In this mode with any unit, when a sample is placed on the pan and
[PRINT]
key is pressed, that value
is output via the RS-232C cable or DATA I/O connect and automatic taring is performed each time
afterwards. This is repeated every subsequent time a new sample is placed and
[PRINT]
key is
pressed. Pressing the
[POWER]
key stops the Formulation mode. When stopped, the measurements
up to that point are summed up and displayed.
When the optional printer or computer is connected, the start up is printed out as “--------- FORMULA-
TION ---------” and upon stop, the total is printed out as “TOTAL=”.
(Setting Formulation Mode)
1
Press the
[CAL]
key repeatedly from the mass
display until “FUnC.SEL” appears. Press the
[O/T]
key. The display will show “CAL”.
2
Press the
[CAL]
key repeatedly until “Formu:
∗∗
”
appears. The “
∗∗
” shows the current setting, “on”
for on, “oF” for off.
3
Press the
[O/T]
key. The display shows “Formu-on”
and after this, pressing the
[CAL]
key switches
the displays between “Formu-on” and “Formu-oF”.
When the current setting is displayed, the stability
mark (
) appears. Press the
[O/T]
key when the
desired setting is shown, to change the setting, or
4
Press the
[POWER]
key to return to “Formu:
∗∗
”.
5
Pressing the
[POWER]
key again returns to the
mass display.
(When OFF)
(When OFF)
Setting
check
(Example)
(Example)
During
setting
(When ON)
(When ON)
During
setting
(Example)
(Example)
Setting
check
Note
No more than one of the four application modes, Auto Print (
), Interval
Timer (
), Add-on Mode (
), Formulation Mode (
) can be set ON
at the same time.