72
13. Application Functions
13.5 Interval Timer
(AUW-D/AUW/AUX series only)
Not applicable to a verified balance as a legal measuring instrument in the EU
This function automatically outputs the measurement values of the balance at set time intervals.
1
Press the
[CAL]
key repeatedly from the
mass display until “FUnC.SEL” appears. Press
the
[O/T]
key to display “CAL”.
2
Press the
[CAL]
key repeatedly until “int”
appears.
3
Press the
[O/T]
key. “int:
∗∗
:
∗∗
” appears.
(
∗∗∗∗
are numbers)
The
symbol and the
symbol appear
in order to show number entry mode.
∗∗
:
∗∗
shows the current interval setting as [MM:SS]
(where MM is minutes and SS is seconds). The
leftmost digit blinks.
4
Pressing the
[UNIT]
key increases the blinking
digit by 1. Pressing the
[PRINT]
key moves the
blinking digit one place to the right. Set minutes
up to 99 and seconds up to 59.
5
Press
[O/T]
key to set.
6
Press the
[POWER]
key repeatedly to return to
the mass display. The letter “T” and STAND-BY
symbol appear indicating interval output standby.
7
Press the
[PRINT]
key. Automatic output com-
mences and continues at the set interval.
8
To pause, press the
[POWER]
key.
9
To discontinue the Interval Timer printing func-
tion, hold down the
[POWER]
key during mass
display for about three seconds until “APL End”
appears. The letter “T” and STAND-BY symbol
will disappear.
(Example)
(Example)
(Example)
(Example)
(When set to 1 minute, 30 seconds)