74
13. Application Functions
13.6 Add-on Mode
This function is convenient for making many measurements of minute samples. This function automati-
cally output the measurement valve and tare after a sample is placed on the pan and the stability mark
displays. Pressing the
[POWER]
key displays and output the total.
Setting
1
Press the
[CAL]
key repeatedly from the mass
display until “FUnC.SEL” appears. Press the
[O/T]
key. The display will show “CAL”.
2
Press the
[CAL]
key repeatedly until “Addon:**”
appears. The “**” shows the current setting, “on”
for on, “oF” for off.
3
Press the
[O/T]
key. The display shows “Addon-
on” and after this, pressing the
[CAL]
key switches
the displays between “Addon-on” and “Addon-oF”.
When the current setting is displayed, the stability
mark (
) appears. Press the
[O/T]
key when the
desired setting is shown, to change the setting.
When the Add-on Mode is set up. “--- ADDON
MODE ---” is outputted.
4
Press the
[POWER]
key to return to “Addon:**”
5
Pressing the
[POWER]
key again returns to the
mass display.
6
When the Add-on Mode is set up, the Add-on
mode symbol and Standby mark appears in the
mass display. It is ready for start.
(Example)
(When ON)
(When OFF)
(When OFF)
(Example)
Setting
check
During
setting
(Example)
(Example)
(Example)
During
setting
Setting
check
(When ON)
(Ready for start)
Note
No more than one of the four application modes, Auto Print (
), Interval Timer (
), Add-on
), Formulation Mode (
), can be set ON at the same time.