OMNI-Test Laboratories, Inc.
Control No. C-SFAK-0001, Effective date: 12/2/2015
Page 1 of 1
DIFFERENTIAL PRESSURE GAUGE CALIBRATION DATA SHEET
Instrument to be calibrated: Pressure Transducer______________
Maximum Range: 0-
2” WC
ID Number: OMNI-00336
Calibration Instrument: Digital Manometer
ID Number: OMNI-00395
Date: 1/17/2018
By:
B. Davis
This form is to be used only in conjunction with Standard Procedure C-SPC.
*Acceptable tolerance is 4%.
The uncertainty of measurement is ±0.4
” WC. This is based on the reference standard having a TAR (Test
Accuracy Ratio) of at least 4:1.
Technician signature:
Date: 1/18/2018
Reviewed by:
Date:
Range of
Calibration Point
(
″WC)
Digital
Manometer
Input
(
″WC)
Pressure Gauge
Response
(
″WC)
Difference
(Input -
Response)
% Error of
Full Span
*
0-20% Max. Range
0 - 0.4
0.35
0.362
0.012
0.6
20-40% Max. Range
0.4 - 0.8
0.65
0.672
0.022
1.1
40-60% Max. Range
0.8
– 1.2
1.00
1.024
0.024
1.2
60-80% Max. Range
1.2
– 1.6
1.30
1.340
0.040
2.0
80-100% Max. Range
1.6
– 2.0
1.70
1.749
0.049
2.45
1/19/2018
47 of 194
47 of 194
Summary of Contents for GF55
Page 29: ...29 of 194 29 of 194...
Page 30: ...30 of 194 30 of 194...
Page 31: ...31 of 194 31 of 194...
Page 32: ...32 of 194 32 of 194...
Page 33: ...33 of 194 33 of 194...
Page 38: ...38 of 194 38 of 194...
Page 39: ...39 of 194 39 of 194...
Page 40: ...40 of 194 40 of 194...
Page 41: ...41 of 194 41 of 194...
Page 42: ...42 of 194 42 of 194...
Page 45: ...1 19 2018 45 of 194 45 of 194...
Page 48: ...1 19 2018 48 of 194 48 of 194...
Page 49: ...49 of 194 49 of 194...
Page 50: ...50 of 194 50 of 194...
Page 52: ...52 of 194 52 of 194...
Page 53: ...53 of 194 53 of 194...
Page 114: ...43 MERIDIAN Components May 2015 4 11 1 17 6 2 12 7 5 3 39 Parts Diagram Components...
Page 117: ...46 Notes...
Page 155: ...36 MERIDIAN CAST IRON Components May 2015 4 11 1 10 6 2 9 7 5 3 8 Parts Diagram Components...
Page 157: ...38 Notes...