34 |
P a g e
OPERATION
S
ET THE
S
HAKER
S
PEED
(RPM)
Perform the steps below to adjust the RPM setpoint.
End of Procedure
1.
Turn on the shaking platform
Verify the chamber lid is completely closed.
Place the RPM switch in the ( I ) ON position.
•
The platform will begin shaking. The platform agitates
continuously as long as the RPM switch is ON.
2.
Navigate to the RPM Setpoint Adjustment mode
Briefly push and release either the
Up
or
Down
arrow
button to access the RPM setpoint adjustment mode.
•
The display will briefly flash the letters “SP”, then
show the flashing, adjustable RPM setpoint.
Note:
The display will automatically exit the adjustment mode after
5 seconds of inactivity, with the last shown setpoint value saved.
Setpoint Adjustment
Mode
Initial Setpoint
3.
Adjust the RPM Setpoint
Use the
Up
and
Down
arrow buttons to adjust the
shaking speed to a value between 30 – 400 RPM.
New Setpoint
4.
Wait 5 seconds after entering the RPM Setpoint
The display will stop flashing, and the setpoint is now
saved in the controller.
•
The shaking speed will now automatically
increase or decrease to match your setpoint.
•
The display will revert to showing the current RPM.
Increasing to Setpoint