32 |
P a g e
OPERATION
Note:
When putting the oven into operation in a new location, set the constant temperature set point
equal to your application or the highest temperature of your multi-step heating profile.
Note:
An active profile overrides the constant temperature set point. The constant temperature set
point cannot be adjusted while a heating profile is active.
S
ET THE
C
ONSTANT
T
EMPERATURE
S
ET
P
OINT
This procedure sets a constant temperature set point. When powered, the oven will heat to and then
maintain the constant temperature set point.
The controller must be on its
Home Page
to set the constant temperature set point.
1.
Use the Up and Down arrow keys to adjust the green constant temperature set point
on the second row to match your set point.
•
Holding down an arrow key causes the temperature to advance in increments
of tens (10) rather ones (1).
2.
Release the Arrow key once you have reached your set point. There may be a brief
pause before the oven starts heating.
Note:
After setting the Over Temperature Limit Setting, the constant temperature set point may be
set to zero (0) or any temperature lower than the present Over Temperature Limit that best facilitates
operations. Note that running the oven at temperature with atmosphere in the chamber—for example
between baking runs—may result in a temperature spike when vacuuming down the chamber.
End of procedure
Natural chamber
resting temperature,
a15°C
Constant
Temperature Set
Point, initially set to
0°C
Constant
Temperature Set
Point 150°C
Heating Indicator
Light “1”
Home Page
Home Page
Profile
Indicator
Summary of Contents for SVAC9-2
Page 1: ...Installation Operation Manual SVAC9 2 Vacuum Oven 230 Voltage...
Page 8: ...8 P a g e INTRODUCTION...
Page 12: ...12 P a g e RECEIVING...
Page 22: ...22 P a g e GRAPHIC SYMBOLS...
Page 26: ...26 P a g e CONTROL OVERVIEW...