USER MAINTENANCE
End of procedure
Temperature Calibration Continued
7
Compare the reference device reading with the chamber temperature
display again.
•
If the reference device and the chamber temperature display
readings are the same or the difference falls within the range of
your protocol,
the incubator is now calibrated for temperature.
-Or-
•
See the next step if the readings fail to match or fall outside of
your protocol range.
Reference Device
8
If the two readings are not the same, and the difference still falls outside
the acceptable range of your protocol, repeat steps 3 – 7 up to two more
times.
•
Three calibration attempts may be required to successfully
calibrate units that are more than ±2°C out of calibration.
Reference Device
9
If the temperature readings of the incubator temperature display and the
reference device still fall outside your protocol after 3 calibration
attempts, contact your distributor or technical support for assistance.
The manufacturer recommends turning the Defrost cycle back on after
completing the calibration.
DEFROST