USER MAINTENANCE
Continued on next page
Temperature Calibration Continued
3
Place the display in temperature calibration mode.
a.
Press and hold both the
UP
and
DOWN
temperature
buttons simultaneously for approximately 5 seconds.
b.
Release the buttons when the temperature display
shows the letters “C O”. The display will begin flashing
the
current temperature display value
.
Note:
If an arrow key is not pressed for 5 seconds, the display will
cease flashing, and store the last displayed number as the new
current chamber temperature value.
4
Use the
Up
and
Down
buttons to adjust the current
display temperature value until it matches the reference
device temperature reading.
Reference Device
5
After matching the display to the reference device, wait 5
seconds.
•
The temperature display will cease flashing and
store the corrected chamber display value.
•
The incubator will now begin heating or cooling in
order to reach the set point with the corrected
display value.
Cooling to Set Point
6
Allow the incubator to sit for at least 1 hour undisturbed to
stabilize
after the incubator has achieved the corrected
temperature set point.
•
Failure to wait until the incubator is fully stabilized
will result in an inaccurate reading.
Set Point Achieved
Summary of Contents for SRI3
Page 1: ...BOD INCUBATOR 110 120 Voltage 220 240 Voltage Installation Operation Manual SRI3 SRI3 2...
Page 8: ...8 P a g e INTRODUCTION...
Page 12: ...12 P a g e RECEIVING YOUR UNIT...
Page 20: ...20 P a g e GRAPHIC SYMBOLS...
Page 22: ...22 P a g e CONTROL PANEL OVERVIEW...
Page 38: ...38 P a g e USER MAINTENANCE...