OUTPUT - Linear in °C
OUTPUT - Linear in ppm(v)
For example:
For a range of -80°C to + 20°C dewpoint the analogue output will be at its
minimum when the reading is at -80°C dewpoint and at its maximum when the
reading is at +20°C dewpoint.
Furthermore the factory default configuration is that the output current ranges
from 4 to 20 mA. Therefore using the above example the port will output 4mA
when the AcuTrak displays -80°C dewpoint.
And 20mA when the AcuTrak di20°C dewpoint.
If a sensor is detected as short circuit then the current output will rise to
+20.25mA.
If required there is facility to change the output current calibration points at either
end of the scale to attain a more focused signal.
8.5.1 Spanning the mA Output
It is possible to change the output current calibration points at either end of the
range to attain a more focused retransmitted signal.
31