Page 21 di 32
PTU50-51-56 -
advanced confi guration
9.2.1 - SET DISTANCE 4mA
Position the cursor on DISTANCE 4mA, press ENTER to access.
Use UP ARROW and SCROLL to modify the value.
Press ENTER to confi rm.
LEFT ARROW to exit without changes
Default value: 1500mm (PTU50 range 1,5mt), 6000mm (PTU51 range 6mt.)
or 12000mm (PTU56 range 12mt)
SET DISTANCE 4mA
SET DISTANCE 20mA
MEDIUM
FILTER COEFFICIENT
BLIND DISTANCE
ACTUAL LEV 4mA
ACTUAL LEV 20mA
Ź
5000 mm
SET DISTANCE 4mA
9.2.2 - SET DISTANCE 20mA
Position the cursor on DISTANCE 20mA, press ENTER to access.
Use UP ARROW and SCROLL to modify the value.
Press ENTER to confi rm.
LEFT ARROW to exit without changes
Default value: 100mm (PTU50 range 1,5mt), 400mm (PTU51 range 6mt.)
or 600mm (PTU56 range 12mt)
0300 mm
SET DISTANCE 20mA
SET DISTANCE 4mA
SET DISTANCE 20mA
MEDIUM
FILTER COEFFICIENT
BLIND DISTANCE
ACTUAL LEV 4mA
ACTUAL LEV 20mA
Ź
9.2.3 - MEDIUM
Position the cursor on MEDIUM, press ENTER to access.
3 confi gurations are possible:
LIQUIDS - liquids measurement
SOLIDS - granular solids measurement
LIQUIDS PIPE - liquids measurement in pipe reference
Press SCROLL to select the product type.
Press ENTER to confi rm.
LEFT ARROW to exit without changes
Default value: LIQUIDS
LIQUIDS
SOLIDS
LIQUIDS PIPE
Ź
SET DISTANCE 4mA
SET DISTANCE 20mA
MEDIUM
FILTER COEFFICIENT
BLIND DISTANCE
ACTUAL LEV 4mA
ACTUAL LEV 20mA
Ź
9.2.4 - FILTER COEFFICIENT
Position the cursor on FILTER COEFFICIENT, press ENTER to access.
Enter a value from 1 to 99: 1 maximum speed, 99 maximum slowness.
The function is deactivated with 0 (immediate response)
Use UP ARROW and SCROLL to modify the value.
Press ENTER to confi rm.
LEFT ARROW to exit without changes
Default value: 10
20
FILTER COEFFICIENT
SET DISTANCE 4mA
SET DISTANCE 20mA
MEDIUM
FILTER COEFFICIENT
BLIND DISTANCE
ACTUAL LEV 4mA
ACTUAL LEV 20mA
Ź