installation
5.1.5 Agitators presence
The level measurement is possible thanks to the Auto-Tuned statistical fi lter.
Should rarely need to adjust the fi lter setting by editing 2 PTU50-51-56 sensor programming parameters:
- FILTER; this parameter is present in the Quick Setup menu and in the Advanced Confi guration “SETUP” menu;
increasing the parameter value, decreases the sensor sensitivity to the level measurement sudden variations.
- F-WINDOW; this parameter is present in the Advanced Confi guration “SERVICE” menu; decreasing the parameter
programmed value, increases the sensor immunity to false echoes.