SI 510 WEIGHING INDICATOR
If you want to set Max capacity as 1,000kg and the division is
0.1 (100g), then just input “1000”.
5-1-2. Test Weight Calibration Mode (Using Test Weight)
STEP 1
1) Remove "CAL-BOLT" on the real panel.
2) Press "CAL-LOCK S/W" inside.
STEP 2 – Setting "Capacity of weighing scale"
1) When "CALIBR" display.
2) Press "F1" key.
3) It will be displayed "CAPA"
4) Input max capacity and press enter key.