SI 550 WEIGHING INDICATOR
61
7-3. Indicator Error & Treatment
Below error is under situation what indicator is not able to measure weight or not able to
progress weighing process.
Display
Cause
Treatment
“
or
“
”
1. Load cell Error
2. Load cell cable Error
3.Load cell connection Error
4. A/D Board Error
5.If Analogue value
is over 1,040,000.
※ When weigh “-“ value,
If it is over set max capa, “OVER”
is displayed.
Ex) Even though set max capa is
“100” and it is over “-100”,
“OVER” is displayed.
1. Under “TEST” mode 1, check
analogue value. If you cannot get
any analogue value or there is no
change although adding load,
please check load cell, load cell
cable, connection conditions first.
2. Replace another load cell, and
check the indicator condition. If
you have same problem, please
replace new
3.
Try to connect the indicator’s A/D
with the other indicator.
4. Check power status
5.
Check the power and connection
of terminal.
“
”
1.
Power is ON, when some
materials are on weighing part
※
Under “Normal Mode”, if there
are more than 10% loading of
Max capacity, “
Un-Pass
” display
will be appeared and indicator
will stay until removing the load.
※
Setting Back-up mode it can
memory empty value, and it
becomes set value without
displaying” Un-pass”)
1.
If you set “Normal Mode”, please
check weighing part empty or not
before turn on the power. If there
are some materials in/on weighing
part, please remove those materials
and turn on the power.
2.
Please try to set F-function 102-
02(Back-up) mode so that the
indicator can remember first empty
value.
“
”
1.
H/W has some problem
.
1.
Please contact the distributor or
Head Office.
※ Under
“CELL-ER”
, Zero key, Tare key, Hold key, and print key will not be activated, and
Analog Output(4~20mA/0~10V), either