6 Maintenance
│
53
6.3
Suitable environment
Device inspection and adjustment must be performed in an envi-
ronment with clean air. This means that the ambient air must be
free of hydrocarbons and toxic gases.
During device inspection and adjustment the device will indicate
whether the ambient air is clean (for instance:
Use clean air!
).
6.4
Device inspection
Device inspections have the task of ensuring the functionality of
the device and, in particular, the display for the measuring mode.
The probes and probe hoses used with the device must also be
visually inspected for each device inspection (/3/).
Note:
In addition to the device inspection, the response time must be
measured and documented regularly.
6.4.1
Ways of carrying out the procedure
The following options are available for carrying out the device
inspection:
●
free
●
guided
Alternatively, the device inspection can be performed using the
test set
ATS 503/501
.
Free device inspection
The user performs the device inspection on his own initiative
and responsibility. The user protocols the results (e.g. on paper).
Guided device inspection
The
PM 580/550/500/400
device leads the user interactive-
ly through all the necessary tests. The tests must be selected
manually.