97
11 Technical aspects
Alarm
Cause, Remedy and Alarm properties
AL1 .............. Lower alarm threshold:
Alarm threshold exceeded
Alarm properties: acoustic alarm clearable, optical
alarm stays on, self-resetting
Exception O
2
sensor (cf. Chap. 3.6):
fallen below the alarm theshold
Alarm properties:clearable when the level of the
alarm threshold exceeds once
again
AL2 .............. Upper alarm threshold:
Alarm threshold exceeded
Alarm properties:clearable when the level falls
once again below the alarm
threshold
AL3 .............. End of measuring range:
Sensor reached end of measuring range
Alarm properties:self-resetting when the level
falls below the end of measuring
range
11.4
Error and Alarm properties
All Fxxx errors do have highest priority i.e. they can only be run
as sole errors.
Errors F40-F44 and F62 are exceptions. These messages can
occur together with alarm messages AL1, AL2 and AL3. The dis-
play then switches between the error and alarm message every
3s. The priority within AL1 to AL3 is from AL1 to AL3 in ascending
order i.e. only an ALx message can be run for a sensor.
Summary of Contents for EX-TEC Combi
Page 1: ...Operating Instructions ...
Page 3: ...Illustration EX TEC Combi EX TEC Combi Device variation 1 Device variation 2 EX TEC Combi 100 ...
Page 5: ...Operating Instructions 15 06 2005 V2 X 103401 en EX TEC Combi ...
Page 117: ...108 Appendix ...
Page 118: ...109 Appendix ...
Page 119: ...110 Appendix ...
Page 120: ...111 Appendix ...
Page 121: ...112 Appendix ...
Page 122: ...113 Appendix ...
Page 123: ...114 Appendix ...
Page 124: ...115 Appendix ...
Page 125: ...116 Appendix ...
Page 126: ...117 Appendix ...