5.2 Displaying alarm settings
Go via MENU/SET UP/DISPLAY to ALARMS - Press ENTER
At display screen
Operation
SELECT COMPONENT
CO
2
vpm
↑
(if more than 1 component)
select the desired component for alarm display, then press
‘ENTER'
[
Measurement
] AL1 ENABLE
10.0 % HI
↑
the first screen for each alarm shows whether it is
enabled
or not
and the
setpoint
and
type
; use the cursor keys to select the next
screen
[
Measurement
] AL1 ENABLE
HYST 1.0% FRZ ;
The second screen for each alarm shows the
hysteresis
and
mode
, use the cursor keys to select the next alarm
[
Measurement
] AL2 ENABLE
10.0 % HI
↑
repeat above process to view all alarm settings, then press ‘EXIT'
5.3 Displaying output configuration
• Relays
Go via MENU/SET UP/DISPLAY/OUTPUTS to RELAYS - Press ENTER
At display screen
Operation
DISPLAY RELAY
1.3 ASSIGNED
↑
Select the desired relay for display, then press ‘ENTER'
1.3
[Measurement]
AL1
HI 10.0%
↑
Display the functions assigned to this relay using the cursor
keys, then press ‘EXIT'
• Analogue outputs
Go via MENU/SET UP/DISPLAY/OUTPUTS to ANALOGUE - Press ENTER
At display screen
Operation
DISPLAY ANALOGUE
1.3 ASSIGNED
↑
Select the desired analogue output for display, then press
‘ENTER’
1.1
[Measurement]
R1
L=[
Lower
] U=[
Upper
]
↑
Display the configuration of this output using the cursor keys,
then press ‘EXIT’