Servomex 2500 Quick Start Manual
18
2500 Series - Configuration and Operation 02500/003E/0
Ethernet enables Modbus TCP communication using Ethernet interface on the
02500913 Communication Board (if fitted)
WARNING
Do not use an uncertified DCS, dilatator, printer or other external equipment in a
Hazardous Area.
3.6 Communication Setup
When the communication mode has been selected as described in section 3.5, it may
be configured using the SET COMMS PARMS menu. Note that this menu is only
displayed if the communication mode is not NONE.
The configuration options displayed are specific to the selected communication mode
and are described in the following sections.
3.7 RS232
The RS232 option uses the following fixed configuration.
BAUD rate:
2400
PARITY:
EVEN
STOP BITS:
1
DATA BITS:
8
HANDSHAKE: HARDWARE
Any device intended for connection to the 2500 must be capable of serial RS232
communication and be configured with these settings.
Entering the RS232 menu displays further options.
SERIAL
This option configures the analyser to transmit serial data frames at user configurable
time intervals. For connection details and signal interpretation, refer to the installation
manual.
Go via MENU/COMMS/SET COMMS PARMS to SERIAL - Press ENTER
At display screen
Operation
FRAME PERIOD =
00s
Select required value using cursor keys, then press ‘ENTER’. A
zero value will disable transmission.
The PRINT MENU is only displayed if the FRAME PERIOD under the SERIAL menu is
set to zero. It is provided to enable the user to download hard copy records of the
2500’s Identity (ID), current history logs (Alarm, Fault and calibration histories) and
complete SET-UP information. Any printer must have at least 80 column width.
Go via MENU/COMMS/SET COMMS PARMS to PRINT - Press ENTER
Summary of Contents for 02500003E
Page 2: ...Servomex 2500 Quick Start Manual 2 2500 Series Configuration and Operation 02500 003E 0...
Page 3: ...Servomex 2500 Quick Start Manual 2500 Series Configuration and Operation 02500 003E 0 3...