Table 1: Specifications
Volume inner buffer chamber
200 ml
Volume outer buffer chamber
450 ml
Voltage (max)
500 V
Current (max)
250 mA
Operating temperature
4 °C – 65 °C
Electrodes
Platinum wire (0.2 mm, 99.99 % pure)
Dimensions
16 x 15.6 x 9.5 cm (WxHxD)
Weight
1.2 kg
3. Operational conditions
This apparatus is intended for indoor use only. Maximum relative humidity up to
80 % (for temperatures up to 31 °C) decreasing linearly to 50 % relative humidity
(for temperatures up to 40 °C), at maximum altitude of 2000 m (MSL).
4. Using the vertical gel electrophoresis unit
4.1. Safety precautions
•
Read
the instructions carefully before using the apparatus.
• Always disconnect the electrophoresis unit from the power supply before removing
the safety cover.
•
Do not
attach the safety lid to the internal gel running unit while it is out of the
bottom buffer chamber.
•
Do not
exceed the maximum operating voltage or current (see Table 1).
•
Do not
operate the electrophoresis units in metal trays.
• Polymerised gels may contain some unpolymerised acrylamide monomer, which is
a volatile, cumulative neurotoxin and suspected carcinogen. Wear effective
protective clothing and gloves. Follow recommended handling and disposal
procedures.
•
Do not
fill the unit with running buffer above the maximum fill lines.
•
Do not
move the unit when it is running.
•
Caution:
During electrophoresis very low quantities of various gases are produced
at the electrodes. The type of gas produced depends on the composition of the
buffer employed. To disperse these gases make sure that the apparatus is run in
a well ventilated area.