16
6.2 Error code
If the instrument shows the error code with beeping sound, press [STOP] button to
stop the beeping sound and press [Enter] button to release of the error status and
make the instrument go to the default setting again.
Error
Possible Causes
Actions
Error 1
RPM Sensor
•
Shut off the power supply, and then, turn on the
power switch again to check the instrument.
•
If the error code shows continuously although
you try to operate again, please contact your
distributor.
Error 2
Door
•
The door is not closed completely.
•
Remove the dirt at the door latch and then close
the door completely again. Check the Door LED
on the display window. If it is not solved in
shortly, open the door with emergency door tool
manually for safety of sample.
Error 3
Motor Overheating
•
The motor is overheated.
•
Shut off the power supply for an hour, and then
turn on the power switch for checking the
instrument.
•
If the error code shows continuously, please
contact your distributor.
Error 4
Low Voltage
•
The power input of Power supply (V / Hz) is
10 % less than required power.
•
Shut off the power supply and then check the
voltage of the Power supply (V / Hz).
•
Use AVR to provide proper power.
Error5
High Voltage
•
The power input of Power supply (V / Hz) is
10 % more than required power.
•
Shut off the power supply and then check the
voltage of the Power supply (V / Hz).
•
Use AVR to provide proper power.
Error 6
Over Speed
•
The instrument is spun with over speed, due to
problems in the overload of motor and the
output of motor.
•
Shut off the power supply, and then, turn on the
power switch again to check the instrument.
Error 7
Software
•
The installed software has bugs.
•
Tuning the firmware (Download)
Error 8
Imbalance
•
Check weight-balances of samples. Turn off and
on the instrument for checking.