5
3 Installation
1. Connect the AC Power cord to the power socket on
the back of the instrument.
2. Turn on the instrument by pressing the switch on the
back of the instrument.
3. For opening the door, press the [DOOR] button.
4. Close the door until hearing a clank shut.
5. When the door is opened, the door LED turns on.
When the door is closed, the door LED turns off.
•
If the door is opened, the instrument could not be operated even with
pressing the [Start] button.
•
For operational safety, this instrument has an automatic rotor recognition
function.
•
When you supply the power, the display shows "Searching Rotor" / “Change
to Rotor ID”. If the rotor is absent, “Error 9” will be displayed. This message
will be cleared after rotor coupling and running.
•
The door is not automatically opened after finishing operation to keep the
sample at proper temperature.
•
Power Failure: If there is any power failure during operation, the door can not
be opened by using the [Door] button. The door can be opened only when
the operation is completely stopped and the power is on again. If you want to
open the door during the power failure, please refer to chapter: “Emergency
Door Open”.
6. Before coupling a rotor, clean the motor shaft and chamber with soft dry towel.