Device installation
AS-11C User Manual
the
"Device Configuration"
tab of the device window. As shown in
"Figure 5.8
Effective scanning range adjustment software operation interface"
, FILPS will
adjust the input value according to the current scanning angle resolution of
AS-11C.
Press the
"Upload to Device"
button to send the configuration data to the AS-11C,
then the AS-11C will automatically restart. After the AS-11C restarts, the newly
set working scan range parameters come into effect. At this time, the AS-11C
only outputs the measurement data within the working scan range, and the area
monitoring function only processes the data within the working scan range.
Related Reading
For details about the adjustment of the scan angle range,
please read "Lidar Diagnostic and Configuration Software
(FILPS) User Manual" Section 7.11 "Specific Angle
Measurement
Data"
and
Section
6.3
"Operation
Configuration Parameters".
Figure 5.7 determines the effective scanning range through measurement data