STC-GE83A
STC-GEC83A
User Guide
25
Rev 1.01
Command No.
Command Description
13H:
MOD4[7..0]
[Camera function mode 4] Initial data: MOD4[7..0] = 60H
Sets the camera function mode.
D[7..0]
D7
D6
D5
D4
D3
D2
D1
D0
D7:
No Function
Always set as “0”
D6:
Trigger signal mask during exposure
0: OFF (No mask)
1: ON (Mask)
D5
Trigger signal mask during image output
0: OFF (No mask)
1: ON (Mask)
D4 to D0:
No Function
Always set as “100000”
Note 1: The trigger signal is invalidated when mask function is on.
16H:
SOFTRG[7..0]
[Software Trigger Setting] Initial data: SOFTRG[7..0] = 80H
Sets the source of the software trigger.
D[7..0]
D7
D6
D5
D4
D3
D2
D1
D0
D7 to D6:
Software trigger source selection
00: Programming software trigger
10: Command software trigger
(200 useconds pulse width trigger signal)
01, 11: No function
(Prohibited settings. Do not set these values)
D5 to D1:
No Function
Always set as “00000”
D0:
Generate software trigger command
0: Hold (Low State)
1: Generate command software trigger
(200 useconds high state)
Note 1: The software trigger source selection is enabled whenever “Trigger signal type (MOD3-D5)” is the software
trigger (set as 0)
Note 2: The “Programming software trigger” is used to set up the pulse duration, trigger signal interval and generate the
trigger signal.
Note 3: When selecting “Command software trigger”, it is necessary to generate the software trigger signal with the
“Generate command software trigger (SOFTRIG-D0)”.
17H:
IMAGEREST
[7..0]
[Image Data Reset] Initial data: IMAGEREST[7..0] = 00H
Reset the Image data (FVAL, LVAL and the image data).
Change from the reset to the image data out after starting the image acquisition.
The image data is not output when resetting the image data.
D[7..0]
D7
D6
D5
D4
D3
D2
D1
D0
D7 to D1:
No Function
Always set as “0000000”
D0:
Image Data Reset
0: FVAL/LVAL/Image data reset
(FVAL, LVAL and the image data are low state data)
1: FVAL/LVAL/Image data out