operational screens will change to instruct
the operator how to silence the Horn.
When the alarm condition is cleared, the
Horn is enabled and will be activated
should another alarm occur.
Clear Alarm:
Silence Horn:
Manual Override:
Manual Override mode
is useful to temporarily measure without
processing alarms. It displays the ppm
level, the refrigerant and the current time.
Alarm processing is suspended during
Manual Override. Manual Override mode is
entered by via the menus described later in
this document.
Menu Screens
The Menu Screens can be accessed from
any Operational Screen by pressing the
"ENT" or enter key. In the first menu, the
operator is given the choice to review Zone
Alarms, manually perform an Auto zero,
enter Manual Override mode or enter
Setup.
Status: T
he Status Screen is accessed by
pressing the "1" key. The status screen will
allow the operator to review alarms and to
review trend data, if trending is enabled on
the monitor.
If trending is not enabled, choice 2)
trending, will not be displayed.
Alarm Status:
The alarm status screen is
accessed by pressing the “1” key. If any
zones are in alarm, the operator will have
access to a series of screens formatted to
show the alarm level, the peak ppm, and
the time and date of the alarm. Additional
alarm screens can be accessed by pressing
the left and right arrow keys. If there are no
alarms to display, or the last alarm has
displayed, "No More Alarms". If the
"RESET" or "ABORT" keys are pressed, the
monitor will exit and return to normal
operation.
Once all of the measurement alarms have
been displayed, the Monitor will check for
Trouble Alarms. Trouble Alarms will be
generated if the monitor senses a fault in
the Infrared Bench (sensor), a fault in the
remote reset connection, a calibration fault
and a flow related failure.
IR Bench Fault:
The Bench Voltage trouble
alarm will be displayed if the monitor
senses a fault in the Infrared Sensor.
Auto Sample AREA 1
PPM 0 12:00
R-134A
RESET to Silence
Auto Sample AREA 1
PPM 0 12:00
R-134A
RESET to Silence
1)…Status
2)…Rezero
3)…Manual
4)…Setup ABORT
Alarm Status
AREA 1 LOW PK 26
3/12/03 12:00
<- RESET or ABORT ->
1)…Alarms
2)…Trending
ABORT