4 Troubleshooting
SenTec Digital Monitoring System
71
Problem
Possible Cause(s)
Recommended action(s)
Thread of the
Service Gas bottle
is damaged
Use a new Service Gas bottle and install it as
described in
Instruction Manual for the SDMS.
SDM is defective
Contact qualified service personnel.
SDM defective
Contact SenTec authorized service personnel.
Docking Station
Module defective
Replace Docking Station Module according to
the
Repair Manual for the SDMS.
To be performed by SenTec authorized
service technicians only
.
P0504
Consumption
of the
Service
Gas is higher
than usual
(message
‘Gas leak in
DS’ does not
display)
Low pressure side
residual gas flow
too high
Verify whether DP01 status code is flagged.
Contact qualified service personnel.
High pressure part
of the Docking
Station Module is
leaky
Contact qualified service personnel.
Unstable sensor
requiring too many
calibrations
Remembrane sensor and observe gas
consumption.
If problem persists, contact qualified service
personnel.
SDM defective
Contact SenTec authorized service personnel.
Gasket in the
Docking Station
Module thread for
the Service Gas is
polluted or
damaged
Clean gasket in the Docking Station Module
thread according to the
Repair Manual for the
SDMS.
To be performed by SenTec authorized
service technicians only
.
Docking Station
Solenoid Valve
defective
If problem persists the Docking Station
Solenoid Valve is defect. Replace it by original
replacement parts according to the
Repair
Manual for the SDMS.
To be performed by SenTec authorized
service technicians only
.